SPECIFICATIONS (PUMP)
|
|
Flow Range |
0.001 — 9.999 Mi/min |
Delivery method |
Double piston ( Main & Auxiliary) |
Pump head |
10m1 SST/Peek analytical |
Accuracy |
± 1% or ± 2pLimin, whichever is greater ± 1% or ± 1pL /min, whichever is greater (Micro)
|
Flow rate settings |
0 - 0.10 mL/min. 0.001 inc. 0.10 - 0.90 mtimin 0.010 inc. 1.0 ml 1.0 mL /Min.+ 0.10 inc
|
Precision |
0.25% from 0.1 mL/min. to 10 mL/min. at 20 deg C 0.25% from 0.05 mL/min. to 4 mL/min. at 20 degrees C (Micro) |
Pmax |
6000psi |
System connections |
1/16" capillaries |
Dimensions |
410mm x 180mm x270mm |
Weight |
7.3kgs |
Power input |
85 to 264 VAC. 47 to 63 Hz, 60 Watts |
Control |
Stand alone mode using front panel |
SPECIFICATIONS (UV DETECTOR)
|
|
Wavelength |
190-700 nm |
Spectral bandwidth |
6 nm |
Wavelength Accuracy |
≤ ± 1 nm |
Wavelength Reproducibility |
≤ ± 0.5 nm |
Flow pool volume |
10 µ L |
Optical path |
10 mm |
Dynamic Noise |
≤2.5x10 AU (Methanol, 1mI/min 254nm) |
Static Noise |
≤ ± 1 <10-5AU (Empty pool. Response time 1 s) |
Dynamic Drifting |
≤ ± 2.5x 10-4 AU (Methanol, 11m1/min, 254nm) |
Static Drifting |
≤ ± lx 10-4 AU (Empty pool, Response time 1 s) |
Qualitative repeatability |
RSD ≤0.1% |
Quantitative repeatability |
RSD 0.5% |
Dimensions |
160H x 260W x 350 D. (mm) |
Weight |
6kg |
Minimum detection concentration |
≤ 1x10-8g/mL (Naphthalene/methanol solution) |